W&T 57126 Ethernet Pt100/Pt1000 thermometer with solid-state relay

W&T 57126
W&T 57126

W&T 57126 Ethernet Pt100/Pt1000 thermometer with solid-state relay


W&T

The W&T Web Thermometer Relay 57126 combines Pt100/Pt1000 temperature measurement, a network data logger and a potential-free solid-state relay output. Data is accessible via the web, OPC UA, MQTT, REST and Modbus TCP. More

Code: 57126

383 EUR excl. VAT (463,43 EUR Incl. VAT)
Delivery 3-8 days
pcs
(0 pcs In stock)

Rezervace termínu

Get a quote
W&T

The W&T Web Thermometer Relay 57126 combines Pt100/Pt1000 temperature measurement, a network data logger and a potential-free solid-state relay output. Data is accessible via the web, OPC UA, MQTT, REST and Modbus TCP. More

Code: 57126

W&T 57126

Ethernet thermometer with solid-state relay output

Price on request

The W&T Web Thermometer Relay 57126 combines Pt100/Pt1000 temperature measurement, a network data logger and a potential-free solid-state relay output. Data is accessible via the web, OPC UA, MQTT, REST and Modbus TCP.

Do you need to select or order the W&T 57126?
Please send us the required quantity and a description of the wiring configuration. We will confirm the price and availability in our quotation.

Enquire about a product and consultation
1× Pt100/Pt1000Measurement input: –200 to +650 °C
Solid-state relayMax. 39 V AC/DC and 11.7 W
PoE or 12–48 V DCNetwork measurement and switching

Usage

It is used to monitor temperatures in technical rooms, switchboards and industrial equipment. The web-based display of data trends and customisable alerts facilitate long-term monitoring without the need for a separate computer to carry out the measurements.

Technical specifications for the W&T 57126

Temperature sensor1x Pt100/Pt1000 measurement input
W&T probe cable: 2 m, can be extended depending on the connection
Probe protection rating: IP67
Digital outputSolid-state relay
Relay output8-pin terminal block
Conductor cross-sectional area 0.2–1.3 mm²
Length of stripped section: 6–8 mm
Network interface10/100BaseT, auto-detection / Auto-MDIX
RJ45
IPv6 on request
Galvanic isolationNetwork interface section: min. 1500 V
Power supplyPower-over-Ethernet (PoE) or
12–48 V DC (±10 %)
Current consumptionPoE Class 1 (0.44–3.84 W)
typ. 69 mA at 24 V DC
max. 80 mA at 24 V DC
Signalling1 Power LED
2× network status LEDs
4 LEDs for status and faults
SensorPt100/Pt1000 connection: two-wire, three-wire or four-wire
Measuring rangeW&T sensor: -50 °C to 180 °C
PT100/PT1000 measurement input: -200 °C to 650 °C
Resolution1/10 °C
Measurement unit error±0.3 °C, ±0.2 %
Class A Pt100/Pt1000 sensor error±0.15 °C, ±0.2 %
Save interval15, 30 s, 1, 5, 15, 60 min
History capacity (6 MB memory)min. 16 weeks, max. 20 years
Internal clock deviationmax. 1 min. / month
Measurement period4 seconds
Digital output1 potential-free contact, solid-state relay
Maximum switching currenttyp. 300 mA AC/DC (peak value 500 mA)
Maximum switching voltage39 V AC/DC
Maximum switching power11.7 W AC/DC
Housing and mountingCompact plastic housing for DIN rail mounting
105 × 22 × 75 mm (length × width × height)
Degree of protectionIP20
Weightapprox. 200 g
Ambient storage temperature-40 to +70 °C
Operating ambient temperature0 to +60 °C
Contents of the delivery1× network thermometer with relay output for DIN rail mounting
1x Pt100 sensor, 2 m, IP67
1× quick start guide

Sensors, ranges and temperature recording

  • The potential-free output consists of a solid-state relay. The specified switching current is typically 300 mA AC/DC, with a peak of 500 mA; the maximum voltage is 39 V AC/DC and the switching power is 11.7 W. This is not a mechanical contact for direct switching of 230 V.
  • One measurement input on the terminal block. The supply includes one Pt100 sensor with a 2 m cable and IP67 protection.
  • The measurement input covers -200 to +650 °C. The supplied W&T sensor is rated for -50 to +180 °C. The usable range of the entire measurement is limited by the specific sensor and its design.
  • Two-wire, three-wire and four-wire Pt100/Pt1000 connections are supported. The selected sensor type and number of wires must correspond to the input settings and wiring.
  • The resolution is 0.1 °C and measurements are updated every 4 seconds. The recording interval can be selected independently: 15 or 30 seconds, or 1, 5, 15 or 60 minutes.
  • Up to 12 customisable alerts are available, along with email, SNMP and other network services. The length of the history retained depends on the number of measured channels and the set interval.

Commissioning and accessories

Select sensors according to the required temperature and environment. Check the wiring, shield connection and Pt100/Pt1000 settings. After setting the IP address, select the units, logging interval and alarm limits. The IP20 rating specified applies to the network module itself, not automatically to the probe.

Contents of the delivery

  • 1× network thermometer with relay output for DIN rail mounting
  • 1x Pt100 sensor, 2 m, IP67
  • 1× quick start guide

Frequently Asked Questions

Does the supplied sensor measure the full range from -200 to +650 °C?

No. This range applies to the measurement input. The supplied W&T sensor has a range of -50 to +180 °C. A suitable, compatible sensor is required for a different range.

How many sensors are included in the delivery?

One measurement input on the terminal block. The supply includes one Pt100 sensor with a 2 m cable and IP67 protection.

Can I switch 230 V directly using the relay output?

No. The solid-state relay output is rated for a maximum of 39 V AC/DC and 11.7 W. Please also take into account current limitations and the nature of the load.

Why buy from FOXON

W&T distributor

We are a distributor of W&T products. We supply new original products with a warranty.

Stock availability and delivery

We check current availability from our own and partner warehouses. Before you order, we confirm the price and possible delivery date for the specific model.

Product selection and technical advice

We help you choose the model, power supply and accessories to suit your connections and communication requirements.

Solution design and configuration

We discuss the connections and application requirements. Where needed, we propose a suitable setup and agree on the scope of assistance with configuration and commissioning.

More than 3,500customers
We supply tomore than 100 countries worldwide
Express delivery?We will check. Contact us now.

Documentation and technical consultancy

Technical data and documentation W&T 57126

Manuals, firmware and tools for this model are also available in the Downloads tab. We can help with product selection and configuration according to your connections and communication requirements.

Do you need to select or order the W&T 57126?
Please send us the required quantity and a description of the wiring configuration. We will confirm the price and availability in our quotation.

Enquire about a product and consultation

Diagrams and illustrations

W&T 57126 – diagram or illustration of the productW&T 57126 – diagram or illustration of the product

Technical specifications for the W&T 57126

Temperature sensor1x Pt100/Pt1000 measurement input
W&T probe cable: 2 m, can be extended depending on the connection
Probe protection rating: IP67
Digital outputSolid-state relay
Relay output8-pin terminal block
Conductor cross-sectional area 0.2–1.3 mm²
Length of stripped section: 6–8 mm
Network interface10/100BaseT, auto-detection / Auto-MDIX
RJ45
IPv6 on request
Galvanic isolationNetwork interface section: min. 1500 V
Power supplyPower-over-Ethernet (PoE) or
12–48 V DC (±10 %)
Current consumptionPoE Class 1 (0.44–3.84 W)
typ. 69 mA at 24 V DC
max. 80 mA at 24 V DC
Signalling1 Power LED
2× network status LEDs
4 LEDs for status and faults
SensorPt100/Pt1000 connection: two-wire, three-wire or four-wire
Measuring rangeW&T sensor: -50 °C to 180 °C
PT100/PT1000 measurement input: -200 °C to 650 °C
Resolution1/10 °C
Measurement unit error±0.3 °C, ±0.2 %
Class A Pt100/Pt1000 sensor error±0.15 °C, ±0.2 %
Save interval15, 30 s, 1, 5, 15, 60 min
History capacity (6 MB memory)min. 16 weeks, max. 20 years
Internal clock deviationmax. 1 min. / month
Measurement period4 seconds
Digital output1 potential-free contact, solid-state relay
Maximum switching currenttyp. 300 mA AC/DC (peak value 500 mA)
Maximum switching voltage39 V AC/DC
Maximum switching power11.7 W AC/DC
Housing and mountingCompact plastic housing for DIN rail mounting
105 × 22 × 75 mm (length × width × height)
Degree of protectionIP20
Weightapprox. 200 g
Ambient storage temperature-40 to +70 °C
Operating ambient temperature0 to +60 °C
Contents of the delivery1× network thermometer with relay output for DIN rail mounting
1x Pt100 sensor, 2 m, IP67